CL-ASI-68
FCL/O3 electrode

NT$100

  • Widely used in drinking water, swimming pools, spas, industrial applications, or dirty/colored samples where colorimetric methods are inadequate.
  • The electric circuit of the whole probe is isolated
  • The interference of monochloramine (a form of combined chlorine) is eliminated so that it can accurately reflect the concentration of Free Chlorine
  • The proprietary Internal Filled Gel (IFG) significantly reduces dependence on pH
  • Excellent stability and long service cycle
  • RS485 communication interface, standard Modbus protocol, easy to integrate

 

 

  • Measuring principle : 3-electrode polarographic method
  • Measuring range : 0~100ppm
  • Accuracy : 5ppm: 4 %
  • pH dependence : If the change of sample pH is less than one unit, pH compensation is not necessary.
  • Response Time : T90 is 1.5~5minutes (depending on the application conditions )
  • Calibration : Usually using one-point calibration with DPD method as reference. Two-point calibration is recommended every 3 to 6 months
  • Communications : RS485 interface, Modbus RTU protocol
  • Cable length : 6m, other lengths can be customized
  • Operating Conditions : 0~50°C, 150 μS/cm wate
  • Power supply : 5V~36VDC, Consumption: average 50mA @ 5V