Home/◉ Sensors/CL-ASI-68FCL/O3 electrode “Waterproof PH/ORP electrode customization” has been added to your cart. View Inquiry List CL-ASI-68FCL/O3 electrodeNT$100 Widely used in drinking water, swimming pools, spas, industrial applications, or dirty/colored samples where colorimetric methods are inadequate. The electric circuit of the whole probe is isolated The interference of monochloramine (a form of combined chlorine) is eliminated so that it can accurately reflect the concentration of Free Chlorine The proprietary Internal Filled Gel (IFG) significantly reduces dependence on pH Excellent stability and long service cycle RS485 communication interface, standard Modbus protocol, easy to integrate CL-ASI-68餘氯/臭氧電極 quantity Add to cart Categories: ◉ Sensors, ◉ FCL/Ozone, FCL/O3 electrode Description Measuring principle : 3-electrode polarographic method Measuring range : 0~100ppm Accuracy : 5ppm: 4 % pH dependence : If the change of sample pH is less than one unit, pH compensation is not necessary. Response Time : T90 is 1.5~5minutes (depending on the application conditions ) Calibration : Usually using one-point calibration with DPD method as reference. Two-point calibration is recommended every 3 to 6 months Communications : RS485 interface, Modbus RTU protocol Cable length : 6m, other lengths can be customized Operating Conditions : 0~50°C, 150 μS/cm wate Power supply : 5V~36VDC, Consumption: average 50mA @ 5V Related products X-FC-003-1Flow cell for residual chlorine/ozone NT$100 ◉ Accessories, ◉ FCL/OzoneAdd to cart X-PB-A60WPWall-mounted waterproof case NT$100 ◉ Accessories, ◉ PH/ORP, ◉ COND/RES, ◉ ION, ◉ Dissolved Oxygen, ◉ FCL/Ozone, ◉ Turbidity/TSS, ◉ CODAdd to cart CW-2XXX 系列工業用全不鏽鋼電導度電極 NT$100 ◉ Sensors, ◉ COND/RES, COND electrodeSelect options This product has multiple variants. The options may be chosen on the product page CW-2XXX-K 系列High temperature resistant clamp two-pole conductivity electrode NT$100 ◉ Sensors, ◉ COND/RES, COND electrodeSelect options This product has multiple variants. The options may be chosen on the product page
Measuring principle : 3-electrode polarographic method Measuring range : 0~100ppm Accuracy : 5ppm: 4 % pH dependence : If the change of sample pH is less than one unit, pH compensation is not necessary. Response Time : T90 is 1.5~5minutes (depending on the application conditions ) Calibration : Usually using one-point calibration with DPD method as reference. Two-point calibration is recommended every 3 to 6 months Communications : RS485 interface, Modbus RTU protocol Cable length : 6m, other lengths can be customized Operating Conditions : 0~50°C, 150 μS/cm wate Power supply : 5V~36VDC, Consumption: average 50mA @ 5V